Range |
Accuracy |
Input
Resistance |
Remarks |
| Function-DCV |
| 400mV |
±(0.3%rdg+2dgt) |
Approx. 100MW |
|
| 4V |
±(0.5%rdg+2dgt) |
Approx. 11MW |
40V
400V
1000V |
Approx. 10MW |
| Function-ACV |
| 400mV |
±(0.8%rdg+5dgt) |
Approx. 100MW |
Accuracy
in the case of sine wave AC
| 400mV range |
40Hz~100Hz |
| 4V range |
40Hz~600Hz |
| 40V range up |
40Hz~1kHz |
|
| 4V |
±(0.5%rdg+5dgt) |
Approx. 11MW |
40V
400V
750V |
±(0.8%rdg+5dgt) |
Approx. 10MW |
| Function-W |
| 400W |
±(0.8%rdg+2dgt) |
Open voltage : approx. 0.4V
The measuring current changes according to the value of resistor to measure.
Accuracy at 400ohm range is measured after cancelIng resistance such as test leads by REL
function. |
| 4kW |
| 40kW |
| 400kW |
| 4MW |
±(1%rdg+2dgt) |
| 40MW |
±(2%rdg+2dgt) |
| Function-Capacity |
| 4nF |
±(7%rdg+5dgt) |
Accuracy
was measured after "0" with REL switch with measuring terminals disconnected. |
| 40nF |
| 400nF |
| 4µF |
| 40µF |
| Function-Hz |
| 99.99Hz |
±(0.3%rdg+3dgt) |
Measuring
range :10Hz~999kHz
Only the highest 3 digits are taken as significant figures in 40.0kHz ~ 99.9kHz in 99.9kHz
range and in 400kHz ~ 999kHz in 999kHz range. |
| 999.9Hz |
| 9.999kHz |
| 99.9kHz |
| 999kHz |
| Function-Hz(LPF) |
| 99.99Hz |
±(0.5%rdg+3dgt) |
Measuring
range : 10Hz~999kHz |
| 999.9Hz |
| Function-DCA |
| 400µA |
±(0.5%rdg+2dgt) |
Approx. 1kW |
|
| 40mA |
Approx. 10W |
| 400mA |
Approx. 1W |
| 10A |
±(1%rdg+2dgt) |
Approx. 0.01W |
| Function-ACA |
| 400µA |
±(1%rdg+5dgt) |
Approx. 1kW |
Accuracy in the
case of sIne wave AC
40Hz~1kHz |
| 40mA |
Approx. 10W |
| 400mA |
Approx. 1W |
| 10A |
±(1.5%rdg+5dgt) |
Approx. 0.01W |
| Function-Continuity
Test |
Open voltage :
approx. 0.4V
Buzzer sounds at approx. 40W |
| Function-Diode
Check |
| Open voltage :
approx. 2.2~3.3V |