| Measuring
Method |
|
| Display Counter |
5000 counts
Bar graph 52 segments max. |
| Range Selection |
Auto and Manual ranges |
| Over Display |
"O.L"mark indication |
| Polarity |
Automatic
selection(only"-"is displayed) |
| Low Battery
Indication |
|
| Sampling Rate |
| Numeral display |
Approx.5times/sec |
| Bar graph |
Approx.60times/sec |
|
| Accuracy
Temperature |
18~28ºC 75% RH max. No condensation |
| Operating
Temperature |
0~50°C 70% RH max. No condensation |
Storage Temperature
(Battery removed) |
-20~60°C 80% RH max. No condensation |
| Environmental
Condition |
Operating altitude < 2000m,
Pollution degree II |
| Power Supply |
6LF22 x 1 |
| Power Consumption |
Approx. 5mA typical |
| Auto Power Off
Timing |
17min |
| Fuse Protection |
0.63A/250V 5x20mm IR 1.5kA
12.5A/500V 6x32mm IR 20kA |
Dimension and
Weight
(including Holster) |
179(H) x 87(W) x 54(D)mm,
approx. 430g |
| Accessories |
Instruction manual x 1
Holster(H-50)x 1
Test lead(TL-82) x 1set
Thormocouple (K-250PC) x 1 |
| Safety |
IEC 1010-1 CAT III Max.600V,
CAT II Max. 1000V |
| E.M.C. |
EN50081-1, EN50082-1 |