Photo:DM508S



These models are compact lightweight simple operation easy-to-use insulation resistance meters having a rigid structure and a new design of modern feeling.
Adoption of inner magnetic field type meter, which excludes influence of external magnetic field for an indicator needle portion, causes a stable indication to be obtained. The Auto discharge function has been installed. These models are in compliance with JIS C1302 1994 and IEC1010-1 CATIII 600V.




1. Equipped with the discharge function
2. Equipped with the MW measuring switch that enables both one-shot and continuous measurement with switch lock function.
3. DC voltage up to 60VDC can be measured.
4. AC voltage up to 600VAC can be measured.
5. Equipped with a strap to suspend from the neck to allow both hands to be used for easurement.




1. Test Lead "TL-508S" 1 set
2. Instruction manual, 1 copy
3. Carrying case "C-08S, 1 piece




AC rectifying method: Half-wave rectification
(Mean value indication rms value converted)
Meter specification: Inner-pole type taut-band system,24A
(PDM508S:48
A)
Battery consumption: Checked by BATTERY CHECK range.
Allowable temperature/humidity: 18 ~ 28șC , 75% RH max., no condensation
Service temperature/humidity: 0 ~ 40șC , 80% RH max., no condensation
Storage temperature/humidity: -10 ~ 50șC , 70% RH max., no condensation
Service ambient condition: Altitude 2000m max., environment pollution 2
Built-in battery: Layer built type alkaline battery 6LR61 x 1 (9V)
Max. power consumption: 2.0W ~ 2.6W at MW range
Continuous service hours*: DM508S Approx. 5.5 hours
DM1008S Approx. 4.0 hours
PDM508S Approx. 4.0 hours
Applicable standard: JIS C 1302-1994 Insulation Resistance Testers
Safety standard: IEC-1010-1 Overvoltage CATIII 600V, protection class 2 (EN61010-1)
Withstand voltage: 3.7kV AC (one minute), between measuring terminal and rear case
Size and weight: H144 x W99 x D43mm, approx. 310g

* when indicating the center scale value of the MW range



MW
(Insulation
Resistance)
MODEL DM508S MODEL DM1008S MODEL PDM508S
Rated measuring voltage Large numeral: 1st effective
Measuring value Small numeral: 2nd effective
500V 0-1-500
-1000M
1000MW
1000V 0-2-1000
-2000M
2000MW
500V 0-0.1-50
-100M
100MW
ACV 0 ~ 600V
DCV 0 ~ 60V
Accuracy MW range
1st effective measurement range: ±5% of reading
2nd effective measurement range: ±10% of reading
0, scale: ±0.7% of scale length
No load voltage: ±30/0 % of raged measuring voltage
Rated measuring current: 1 ~ 1.2mA
Short Circuit current: max 2mA

Home     |     About Us     |    Products     |    Telecom Systems    |    Contact Us    |    E-mail


Bhumika Enterprises
1301  Dundas Street West, Suite # 504,
Mississauga, Ontario, L5C 1C9, Canada

Ph: +1(
647) 273 1281
FAX: +1(905) 615 9646
E-mail: info@bhumika.ca
Website: http://www.bhumika.ca

Technical Specifications are subject to change without notice.
Windows is the registered Trademark of Microsoft Corporation, USA.
All other trademarks are the ownership of their respective owners.